Inappropriate S-ICD patient receives false positive shocks

KORPAS, David

Basic information

Original name

Inappropriate S-ICD patient receives false positive shocks

Authors

KORPAS, David (203 Czech Republic, guarantor, belonging to the institution)

Edition

Žilina, Trendy v biomedicínskou inženierstve 2019, p. 1-3, 3 pp. 2019

Publisher

Žilinská univerzita

Other information

Language

English

Type of outcome

Proceedings paper

Field of Study

20601 Medical engineering

Country of publisher

Slovakia

Confidentiality degree

is not subject to a state or trade secret

Publication form

storage medium (CD, DVD, flash disk)

References:

Organization

Fakulta veřejných politik v Opavě – Slezská univerzita v Opavě – Repository

ISBN

978-80-554-1587-1

Keywords in English

false-positive therapy; defibrillation; implantable cardioverter-defibrillator
Changed: 17/9/2019 21:39, David Korpas

Abstract

In the original language

Subcutaneous implantable cardioverter defibrillator (S-ICD) protects the patients at risk for sudden cardiac death while leaving the heart and vasculature untouched. It provides life-saving therapy, but may also deliver inappropriate therapy. Presented case demonstrates a possibility of S-ICD therapy induction due to double-counting. It was original caused by lack of suitable sensing vectors and the solution was possible just particularly.